Utility of the Platelet Function Analyzer in Patients with Suspected Platelet Function Disorders: Diagnostic Accuracy Study.

Fiche du document

Discipline
Type de document
Périmètre
Langue
Identifiants
Relations

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/doi/10.1055/s-0040-1721502

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/pmid/33376942

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/eissn/2512-9465

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/urn/urn:nbn:ch:serval-BIB_03605D5075AF5

Licences

info:eu-repo/semantics/openAccess , CC BY 4.0 , https://creativecommons.org/licenses/by/4.0/




Citer ce document

J. Kaufmann et al., « Utility of the Platelet Function Analyzer in Patients with Suspected Platelet Function Disorders: Diagnostic Accuracy Study. », Serveur académique Lausannois, ID : 10.1055/s-0040-1721502


Métriques


Partage / Export

Résumé 0

Introduction The platelet function analyzer (PFA) is widely used as a screening tool for bleeding disorders in various settings. The diagnostic performance regarding platelet function disorders (PFDs), which are among the most common inherited bleeding disorders, is however still elusive. We aimed to assess the diagnostic value of PFA for PFD in clinical practice. Methods Comprehensive clinical and laboratory data of all consecutive patients referred to a specialized outpatient between January 2012 and March 2017 with a suspected bleeding disorder were prospectively recorded. The diagnostic work-up was performed according to a prespecified protocol following current guidelines and platelet function was tested using light transmission aggregometry as well as flow cytometry. Results Five hundred and fifty-five patients were included (median age 43.7 years; interquartile range [IQR] 29.3, 61.7; 66.9% female). Possible PFD was diagnosed in 64 patients (11.5%) and confirmed PFD in 54 patients (9.7%). In patients with confirmed PFD, median closure times were 107 seconds (ADP or adenosine diphosphate; IQR 89, 130) and 169 seconds (EPI; IQR 121, 211). In patients without bleeding disorders, PFA closure times were 96 seconds (ADP; IQR 83, 109) and 137 seconds (EPI; IQR 116, 158). The sensitivity was 19.5% in case of PFA ADP (95%CI 12.6, 30.0; specificity 86.4%; 95% CI 82.4, 89.8), and 44.3% in case of PFA EPI (95% CI 34.9, 53.9; specificity 75.6%; 95% CI 70.8, 79.9). Conclusion The diagnostic performance of PFA for PFD was moderate to poor. Our results do not support the utilization of PFA for screening of PFD in clinical practice.