The influence of cardiac arrhythmias on the detection of heartbeats in the photoplethysmogram: benchmarking open-source algorithms.

Fiche du document

Date

14 février 2024

Type de document
Périmètre
Langue
Identifiants
Relations

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/doi/10.1088/1361-6579/ad2216

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/pmid/38266291

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/eissn/1361-6579

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/urn/urn:nbn:ch:serval-BIB_154131911BD91

Licences

info:eu-repo/semantics/openAccess , CC BY 4.0 , https://creativecommons.org/licenses/by/4.0/




Citer ce document

L. Jeanningros et al., « The influence of cardiac arrhythmias on the detection of heartbeats in the photoplethysmogram: benchmarking open-source algorithms. », Serveur académique Lausannois, ID : 10.1088/1361-6579/ad2216


Métriques


Partage / Export

Résumé 0

Objective.Cardiac arrhythmias are a leading cause of mortality worldwide. Wearable devices based on photoplethysmography give the opportunity to screen large populations, hence allowing for an earlier detection of pathological rhythms that might reduce the risks of complications and medical costs. While most of beat detection algorithms have been evaluated on normal sinus rhythm or atrial fibrillation recordings, the performance of these algorithms in patients with other cardiac arrhythmias, such as ventricular tachycardia or bigeminy, remain unknown to date.Approach. ThePPG-beatsopen-source framework, developed by Charlton and colleagues, evaluates the performance of the beat detectors namedQPPG,MSPTDandABDamong others. We applied thePPG-beatsframework on two newly acquired datasets, one containing seven different types of cardiac arrhythmia in hospital settings, and another dataset including two cardiac arrhythmias in ambulatory settings.Main Results. In a clinical setting, theQPPGbeat detector performed best on atrial fibrillation (with a medianF 1 score of 94.4%), atrial flutter (95.2%), atrial tachycardia (87.0%), sinus rhythm (97.7%), ventricular tachycardia (83.9%) and was ranked 2nd for bigeminy (75.7%) behindABDdetector (76.1%). In an ambulatory setting, theMSPTDbeat detector performed best on normal sinus rhythm (94.6%), and theQPPGdetector on atrial fibrillation (91.6%) and bigeminy (80.0%).Significance. Overall, the PPG beat detectorsQPPG,MSPTDandABDconsistently achieved higher performances than other detectors. However, the detection of beats from wrist-PPG signals is compromised in presence of bigeminy or ventricular tachycardia.