An investigation of aspects of radiochemical purity of 99mTc-labelled human serum albumin nanocolloid.

Fiche du document

Date

11 octobre 2021

Type de document
Périmètre
Langue
Identifiants
Relations

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/doi/10.1186/s41181-021-00147-8

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/pmid/34633572

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/eissn/2365-421X

Ce document est lié à :
info:eu-repo/semantics/altIdentifier/urn/urn:nbn:ch:serval-BIB_55DDE848CEBB0

Licences

info:eu-repo/semantics/openAccess , CC BY 4.0 , https://creativecommons.org/licenses/by/4.0/




Citer ce document

R. Cusnir et al., « An investigation of aspects of radiochemical purity of 99mTc-labelled human serum albumin nanocolloid. », Serveur académique Lausannois, ID : 10.1186/s41181-021-00147-8


Métriques


Partage / Export

Résumé 0

Nanocolloidal human serum albumin radiolabelled with 99m Tc provides a diagnostic radiopharmaceutical for sentinel node lymphoscintigraphy. NanoHSA (Nanotop), a commercially available kit, enables the simple preparation of this radiopharmaceutical via reconstitution with pertechnetate eluted from a generator. Thin-layer chromatography is widely used for determining radiochemical purity in clinical nuclear medicine. Quality control methods recommended by the manufacturer were sometimes reported to yield variable results. Therefore, we proposed and evaluated three alternative thin-layer chromatography methods for the quality control of [ 99m Tc]Tc-NanoHSA from a commercially available kit. The radiochemical purity of [ 99m Tc]Tc-NanoHSA determined with all methods was reproducible and met the requirements of the SPC and the European Pharmacopoeia (≥ 95%). Our quality control using iTLC-SG chromatographic paper in methyl ethyl ketone mobile phase identified only free pertechnetate as impurity, resulting in > 99% RCP. The quality control using iTLC-SG in 85% methanol or iTLC-SA in 0.9% NaCl identified an additional small fraction of a hydrophilic impurity, resulting in 95-97% RCP. Glucose was identified as a potential 99m Tc-carrying hydrophilic species contributing to hydrophilic impurities. Our quality control of [ 99m Tc]Tc-NanoHSA with non-polar mobile phase tended to underestimate the amount of hydrophilic impurities, although without compromising the final quality of the radiopharmaceutical. Alternative TLC methods using aqueous mobile phases enabled a more accurate determination of hydrophilic impurities.