Fast determination of glucosamine in pharmaceutical formulations by high performance liquid chromatography without pre-column derivatization

Fiche du document

Date

2014

Type de document
Périmètre
Langue
Identifiants
Relations

Ce document est lié à :
http://www.redalyc.org/revista.oa

Licence

Acta Universitaria




Citer ce document

Armando Alcázar Magaña et al., « Fast determination of glucosamine in pharmaceutical formulations by high performance liquid chromatography without pre-column derivatization », Acta Universitaria, ID : 10670/1.pcqrwl


Métriques


Partage / Export

Résumé 0

"The goal of this work was to achieve fast determination of glucosamine in pharmaceutical for - mulations by high performance liquid chromatography. Commercial tablets were dissolved in water and after filtration, 6 μL were injected to Luna C18 column (250 mm × 4.6 mm, 5μm). Diode Array Detector (DAD) was set at 193 nm. Using sodium perchlorate (50 mM, pH 6.5): acetonitrile (99:1, v/v) at 0.8 mL min -1 as a mobile phase, glucosamine eluted with the retention time 2.834 min ± 0.003 min and the peak purity in real sample was 998.9 ± 3.3. The detection limits evaluated for peak height and peak area measurements were 0.05 μg mL -1 and 0.02 μg mL -1 , respectively. The between-day precision relative standard deviation (RSD) evaluated at two concentration levels was always below 1%."